The ET5410 and ET5411 are single-channel DC electronic loads designed for a variety of testing applications, including chargers, switching power supplies, linear power supplies, and batteries. They offer high resolution and precision, with 1mV/10mV and 1mA/10mA capabilities, and come equipped with 12 common modes and comprehensive testing functions.
Function Description
The ET54 series electronic loads provide a range of basic measurement modes:
- Constant Current (CC) Mode: The load consumes a constant current regardless of input voltage changes.
- Constant Voltage (CV) Mode: The load draws sufficient current to maintain the input voltage at a set level.
- Constant Resistance (CR) Mode: The load acts as a constant resistance, consuming current that varies with the input voltage.
- Constant Power (CP) Mode: The load maintains a constant power consumption by adjusting the current as the input voltage changes.
- Constant Current and Constant Voltage (CC+CV) Mode: Combines constant current and constant voltage characteristics.
- Constant Resistance and Constant Voltage (CR+CV) Mode: Combines constant resistance and constant voltage characteristics.
Beyond basic modes, the devices offer specialized testing functions:
- Battery Test: Used to evaluate battery discharge performance in constant current or resistance mode, determining discharge capacity.
- LED Test: Simulates real LED lamp characteristics by incorporating a diode on-voltage setting in CR mode. The load begins to operate when the input voltage exceeds the diode on-voltage, accurately reflecting the load capacity of LED driving power supplies.
- Dynamic Test Mode: Allows repeated switching between two load setting currents or voltages to test the dynamic characteristics of a power supply. This mode supports continuous, pulse, and trigger operations.
- Continuous Mode: The load continuously switches between A and B values after the test starts.
- Pulse Mode: For each trigger signal received, the load switches from A to B, maintains the pulse duration of B, and then switches back to A.
- Trigger Mode: The load switches between A and B values for each trigger signal received, with pulse width settings being ineffective in this mode.
- Scanning Test Mode: Detects continuous operation of the source within a certain range, capturing critical parameters like protection current and turning voltage. Users can set start/end points, step quantity, delay, and threshold types.
- List Test Mode: Tests the source under different loading conditions by pre-setting a sequence of steps, including load mode, load size, delay time, and comparison limits. This is useful for automated production line testing.
- Short Circuit Test: Simulates a short-circuit at the input, with the actual short-circuit value depending on the power supply's maximum output.
- Qualification Test: An additional function for basic measurement modes (CC/CV/CR/CP) that detects in real-time whether the current test is within set limits, displaying "Pass" or "Fail."
Important Technical Specifications
Model Specifics:
- ET5410: Single channel, 400W power.
- ET5411: Single channel, 400W power.
- ET5420: Double channel, 400W power.
Rated Input:
- Input Voltage: 0-150V (ET5410, ET5420), 0-500V (ET5411).
- Input Current: 0-40A (ET5410), 0-40A (Each channel 20A for ET5420), 0-15A (ET5411).
Constant Voltage Mode:
- Range: 0.1-19.999V, 0.1-150.00V (ET5410, ET5420); 0.1-19.999V, 0.1-500.00V (ET5411).
- Resolving Power: 1mV, 10mV.
- Accuracy: ±(0.05%+0.02%FS).
Constant Current Mode:
- Range: 0-3.000A, 0-40.00A (ET5410); 0-3.000A, 0-20.00A (ET5420); 0-3.000A, 0-15.00A (ET5411).
- Resolving Power: 1mA, 10mA.
- Accuracy: ±(0.05%+0.05%FS) for 0-30A (ET5410, ET5420, ET5411); ±(0.08%+0.1%FS) for 30-40A (ET5410).
Constant Resistance Mode:
- Range: 0.05-4.5KΩ.
- Resolving Power: 16bit.
- Accuracy: ±(0.1%+0.5%FS).
Constant Power Mode:
- Range: 0-400W (ET5410, ET5411); 0-200W (ET5420).
- Resolving Power: 10mW.
- Accuracy: ±(0.1%+0.5%FS).
Dynamic Test Function:
- Pattern: CC.
- T1&T2: 1ms-3600s; Resolving power: 1ms.
- Accuracy: 0.1%+1mS.
Battery Test Function:
- Voltage Range: 0-120V.
- Discharge Capacity: 240Ah.
- Test Time: 1s-16h.
Readback Values (Voltage, Current, Power):
- Voltage: Range 0-19.999V, 0-150.00V (ET5410, ET5420); 0-19.999V, 0-500.00V (ET5411). Resolving power 1mV, 10mV. Accuracy ±(0.05%+0.1%FS).
- Current: Range 0-3.000A, 0-40.00A (ET5410); 0-3.000A, 0-20.00A (ET5420); 0-3.000A, 0-15.00A (ET5411). Resolving power 1mA, 10mA. Accuracy ±(0.05%+0.1%FS) for 0-30A; ±(0.08%+0.1%FS) for 30-40A.
- Power: Range 400W (ET5410, ET5411); 200W (ET5420). Resolving power 10mW. Accuracy ±(0.1%+0.5%FS).
Scope of Protection:
- Overvoltage Protection: Greater than 155V (ET5410, ET5420); greater than 520V (ET5411).
- Overcurrent Protection: More than 45A cut off input (ET5410); more than 22A cut off input (ET5420); more than 16A cut off input (ET5411).
- Over Power Protection: 420W (ET5410, ET5411); 220W (ET5420).
- Overtemperature Protection: 85°C.
Usage Features
- User-Friendly Design: Features a 2.8-inch TFT LCD display with rich content, supporting both Chinese and English. The intuitive interface display system ensures simple and convenient operation.
- Key Lock Function: Prevents misoperation by locking keys.
- External Keyboard Support: Equipped with a PS2 interface to support an external keyboard for setting data values.
- Remote Operation: Can be operated remotely via USB and RS-232 (optional) interfaces, allowing control through host computer instructions.
- Trigger Function: Supports manual triggering (front panel [Trigger] button), external triggering (rear panel trigger port), and bus triggering (program-controlled instruction via RS-232).
- Data Storage: Maintains data storage even after power-off.
- Buzzer Alarm: Built-in buzzer for alarms.
Maintenance Features
- Safety Grounding: Ensures reliable connection of the power line's grounding terminal to the protective grounding terminal for personal safety.
- Suitable Fuse: Requires specified type and rated fuse for continuous fire protection.
- Abnormality Handling: If the product is abnormal, it should not be used. Protective measures may be damaged, and unauthorized adjustments are discouraged. The product should be returned for repair or professional inspection.
- Environmental Considerations: Conforms to WEEE directive (2002/96/EC), indicating that electronic/electrical equipment should not be disposed of with household waste. It is recommended to recycle the product to avoid releasing harmful substances.
- Multiple Security Protection: Includes over-voltage, over-current, over-power, and over-temperature protection, with flexible parameter settings to safeguard the load.
- Intelligent Fan Speed Control: Effectively reduces fan noise during operation.
- Reverse Polarity Protection: Protects against incorrect input polarity.