2.11 Analogue outputs
4 to 20mA – programmable as follows:
Range Gas measured Low End High End
Option 1
Range 1
H
2
4mA equiv. to 0%
20mA equiv. to 1.00%…5.00%
- user adjustable
Range 2
Cl
2
4mA equiv. to 0% 20mA equiv. to 100% - fixed
Option 2
Range 1
H
2
4mA equiv. to 0%
20mA equiv. to 2.00%…10.00%
- user adjustable
Range 2
Cl
2
4mA equiv. to 0% 20mA equiv. to 100% - fixed
Maximum load resistance on outputs: 600 ohms
2.12 Fault indicator outputs
Two fault-indicating relay outputs are provided:
1. Instrument status
2. Reactor status ±10°C of set point.
Both are volt-free SPDT relay contacts rated at 30V AC or DC, 0.5A. The “no-fault” condition
is NC - see Figure 6.
2.13 Concentration alarms
Two user-configurable Alarm outputs are provided:
1. Programmable alarm for Cl
2
2. Programmable alarm for H
2
Both are volt-free SPDT relay contacts rated at 30V AC or DC, 0.5A. The “no-alarm”
condition is NC - see Figure 6.
2.14 Speed of response
T90: 30 seconds typical
2.15 Power requirements
Voltage: 99...132V or 198…264V AC, 50/60Hz
Power: 80VA
2.16 Electronics unit
Mounting: Wall mounting case
Dimensions: 650mm (H) x 450mm (W) x 250mm (D) - without glands
Materials:
Polyester enclosure -
(IP66 rating when door is closed and with suitable cable glands)
Net weight: 20 kg (nom.)
2.17 Katharometer/reactor unit
Mounting: Wall mounting
Dimensions: 150mm (h) x 240mm (w) x 115mm (d) including heat shield
Katharometer
Cable:
2 metre length provided -
cable must not be extended or certification is affected.
Net weight: 3 kg (nom.)