EMC guidance and manufacturer’s declarations –Electromagnetic immunity – For non-life-support
equipment
EMC guidance and manufacturer’s declarations –Electromagnetic immunity
EM-1201/1202/1203 ECG Machine are intended for use in the electromagnetic environment specified
below. The customer or the user of EM-1201/1202/1203 ECG Machine should assure that it is used in such
and environment.
Immunity test
IEC 60601
test level
Compliance
level
Electromagnetic environment - guidance
Conducted RF
IEC61000-4-6
Radiated RF
IEC61000-4-3
3Vrms
150kHz-80MHz
Outside ISM band
a
10Vrms
150kHz-80MHz
In ISM band
a
10V/m
80 MHz to 2.5 GHz
3V(V
1
)
10V(V
2
)
10V/m (E
1
)
Portable and mobile RF communications
equipment should be used no closer to any
part of EM-1201/1202/1203 ECG Machine,
including cables, than the recommended
separation distance calculated from the
equation applicable to the frequency of the
transmitter.
Recommended separation distance
p
1
5.3
⎥
⎦
⎤
⎢
⎣
⎡
=
v
d
p
2
12
⎥
⎦
⎤
⎢
⎣
⎡
=
v
d
p
E1
2.1
⎥
⎦
⎤
⎢
⎣
⎡
=d
80MHz to 800MHz