USER MANUAL
MINIVAC
MICRODERMABRASION AND VACUUM THERAPY
ABOUT THIS MANUAL
Version: 2019.11.01
Year: 2019
Language: English
All the rights to this manual are reserved, especially reproduction, publication and translation rights. No portion of this manual may be
reproduced by no means (photocopy, microfilm or other methods) without previous written authorization from ECLERIS. Neither can it
be reproduced, multiplied or published through any electronic data processing form.
The information contained in this manual may be modified or broaden without previous notification, without it representing any
responsibility to ECLERIS.
ECLERIS S.R.L.
FRANCISCO N. LAPRIDA 4955. VILLA
MARTELLI
(B1603ABK) BUENOS AIRES.
ARGENTINA
Tel/Fax: +54 11 48383200
Web: www.ecleris.com
E-Mail: info@ecleris.com
Directora Técnica: Farm. Doris Inés Rasi
Sanchez, MN: 12.856
Autorizado por la ANMAT PM-1120-43
Condición de venta: “Venta exclusiva a
profesionales e instituciones sanitarias”.

Questions and Answers

  • J
    James MccallAug 21, 2025
    How to troubleshoot an Ecleris Medical Equipment Minivac that does not start?
    • N
      Nicholas ClarkAug 21, 2025
      If your Ecleris Minivac isn't starting, consider these possible causes: * It might not be properly connected to the electrical network. Ensure the power supply cable is securely connected. * The power supply cable could be damaged. If it's broken or cut, replace it. * The plug might be in poor condition. If so, replace the plug. * The fuse might be blown. Replace the fuse.
  • A
    Andrew CollierAug 31, 2025
    Why is my Ecleris Medical Equipment Minivac making a strange noise?
    • S
      Stanley HunterAug 31, 2025
      If your Ecleris Minivac is making unusual noises, there are a couple of potential reasons: * The voltage selection could be incorrect. Make sure the voltage selector switch matches the voltage of your electrical network. * The device may not be on a stable surface. Place the device on a vibration-free surface.
  • E
    Ernest PerezNov 10, 2025
    What to do if my Ecleris Medical Equipment Minivac starts but doesn't produce vacuum?
    • J
      josephhurstNov 11, 2025
      If your Ecleris Minivac starts but isn't creating a vacuum, several factors could be at play: * The vacuum regulator might be completely open. Try closing it by turning it clockwise. * The suction piping could be throttled. Ensure the hose isn't pinched or blocked. * There might be leakages in the piping. Inspect the piping and replace any damaged sections. * The collecting bottle may not be sealed correctly. Check that the bottle is properly closed and the O-ring is in place. * The filter could be obstructed. Remove it from the bottle and clean it thoroughly. * The voltage selector switch is set at 220V and the electrical network in use is of 110 V. Place the voltage selector switch according to the electrical network in use. * There could be dirt or debris inside the hose ...

Ecleris MINIVAC Specifications

General IconGeneral
CategoryMedical Equipment
Noise Level<60 dB
Weight5kg
Power Supply50/60 Hz
Dimensions300 x 200 x 300 mm
Foot pedalOptional