SD5/SD6 Ultrasonic TableTop Doppler User Manual Safety Guide
- 2 -
WARNING
7 Do not touch the signal input/output connector and the patient simultaneously.
8 Do not apply this device and other ultrasonic equipment simultaneously on the same
patient, in case of possible hazard caused by leakage current superposition.
9 Do not apply this device simultaneously with other PATIENT-connected equipment,
such as, a cardiac pacemaker or other electrical stimulators, on the same patient.
10 The device is not protected against defibrillation.
11 Do not use the device with HF surgical equipment.
12 Only use the probes provided by the manufacturer.
13 Before using the battery, make sure to read the user manual and safety precautions
thoroughly.
14 Do not heat or throw the battery into fire as this may cause explosion.
15 Do not solder the leading wire and the battery terminal directly.
16 Do not destroy the battery: Do not pierce the battery with a sharp object such as a
needle; do not hit with a hammer, step on or throw or drop to cause strong shock; do
not disassemble or modify the battery.
17 The battery should be charged, used or stored away from the static electricity.
18 Do not mix the battery with metal objects to avoid short-circuit.
19 The battery should be charged at least once a month to prevent overdischarge.
20 Replacement or charging of the battery shall be done at least 1.5 meters away from
patients.
21 The device shall only be used when the battery cover is closed.
22 If the device is not used for a long time, please remove the battery and store it as
required.
23 If the liquid leak from the battery spills onto your skin or clothes, wash well with fresh
water immediately.
24 If the liquid leak from the battery gets into eyes, do not rub the eyes. Wash them
well with clean water and see a doctor immediately.
25 Stop using the battery if abnormal heat, odor, discoloration, deformation or any other
abnormal condition is detected during use, charge, or storage. Please dispose it
according to the local regulations.
26 Do not immerse, throw, or wet the battery in water/seawater.
27 Do not connect any equipment or accessories that are not approved by the
manufacturer or that are not IEC 60601-1 approved to the device. The operation or
use of non-approved equipment or accessories with the device is not tested or
supported, and device operation and safety are not guaranteed.