7. If this occurs, reduce the solvent o
w rate with the manually adjustable valve on
the owmeter unl the motor current is less than maximum allowable motor
current and/or the noise level of the pump has reduced.
8. Repeat the ming test to determine the permissible solvent ush ow rate,
without exceeding the allowable motor current for a specic duraon (refer the
datasheet of the motor protecon switch).
9. Once the ow rate is set, it does not need to be adjusted again and the solenoid
valve can be used to start and stop the quanty of solvent required.
10. During the solvent ush procedure, we recommend you to monitor the state of the
solvent leaving the exhaust of the pump. Make sure that most of the solvent
leaving the exhaust is in the liquid phase.
05/2021 - ©Edwards Limited
Page 18A41892880_C
A41892880_C - Commissioning