USER MANUAL RETEVET™ PORTABLE ERG
T +49 7461 96 580 0 | F +49 7461 96 580 90 | export@eickemeyer.com | www.eickemeyer.com
Rat, mouse scotopic 3 step left eye
Description Eye Flash Luminance Energy
(0.33, 0.33 White)
Background Luminance
(0.33, 0.33 White)
# Flashes
Dark-adapted 0.002 Right 0.002 cd·s/m² @ 0.5 Hz Off 3
Dark-adapted 8.0 Right 8.0 cd·s/m² @ 0.1 Hz Off 1
Rat, mouse scotopic 3 step right eye
Description Eye Flash Luminance Energy
(0.33, 0.33 White)
Background Luminance
(0.33, 0.33 White)
# Flashes
Dark-adapted 0.002 Left 0.002 cd·s/m² @ 0.5 Hz Off 3
Dark-adapted 8.0 Left 8.0 cd·s/m² @ 0.1 Hz Off 1
Other Protocols
The RETevet™ device has two protocols for which measurements are not obtained: The other two protocols are “fl ashlight”
protocols where the device creates either 30 cd/m² or 300 cd/m² white light.
Custom Protocols
If there is a protocol that you would like to run that is not built in the RETevet™ device, a custom protocol can be created for
you. Contact EICKEMEYER® for more information on how to obtain and use a custom protocol. Typically, custom protocols
should defi ne, which tests will be run and what will be the conditions of those tests such as the stimuli intensity, frequency,
color, duration, and type (on-off, ramp, or sinusoidal stimuli).
Custom protocols must be transferred in the Protocols folder on the device. The built-in protocols can be viewed on the device
in the folder EMR / Built-in protocols, which can be a starting point for creating your own custom protocols. Protocols are
written in the full-featured Lua programming language.
Reference Intervals
The RETevet™ device has built-in infrastructure to handle and report your reference interval data. We doe not provide non-
human reference data, as there are dependencies on species, breed, and electrode type used. However, if in your practice you
have tested many subjects with normal vision, contact EICKEMEYER® support for assistance in incorporating your own
reference data.