Instruction Manual
for the
Aquameter
TM
Series
Multiparameter Water Quality
Meters, Probes, Utilities & Accessories
(Software Version 3.00 and above including Optical DO & Depth Measurement)
Document No. 10401-00188
Revision: M
Date: February 16, 2011
Aquaread Ltd
Kent Innovation Centre
Thanet Reach Business Park
Millennium Way
Broadstairs
Kent
CT10 2QQ
ENGLAND
Phone: +44 1843 609 692
www.aquaread.co.uk
P.O. Box 4, 6987 ZG Giesbeek
Nijverheidsstraat 30,
6987 EM Giesbeek,
The Netherlands
T +31 313 880 200
E info@eijkelkamp.com
I www.eijkelkamp.com
M185701E

Table of Contents

Questions and Answers

  • T
    timothy87Aug 21, 2025
    Why are dissolved oxygen readings inaccurate on EIJKELKAMP Measuring Instruments?
    • M
      Melanie DayAug 21, 2025
      Inaccurate or unstable dissolved oxygen readings on your EIJKELKAMP Measuring Instruments can stem from several causes: * There may be insufficient flow across the Galvanic DO membrane. Ensure a constant water flow whilst taking readings (does not apply to AP-900 & AP-1000). * The DO electrode may need calibrating, so recalibrate it. * The DO membrane may be dirty, so clean it. * The DO membrane may be worn out or damaged, so replace the DO membrane cap. * Calibration may have been carried out at an extreme temperature. Recalibrate at a temperature as close to the sample temperature as possible.
  • V
    veronica72Aug 23, 2025
    Why are EC readings inaccurate on EIJKELKAMP Measuring Instruments?
    • E
      Eric MasonAug 23, 2025
      Inaccurate or unstable EC readings, or Error 09 during calibration on your EIJKELKAMP Measuring Instruments, can be due to: * Missing Probe Sleeve: EC will not work without the whole Probe Sleeve fitted. * Insufficient sample depth: Ensure the sample level reaches the minimum depth line on the outside of the Aquaprobe. * Trapped air bubbles: Tap and swish the Aquaprobe to dislodge them. * Loose Probe Sleeve: The Probe Sleeve must be absolutely rigid with respect to the Probe Body for correct EC operation. If you can move the Probe Sleeve to and fro whilst holding the Probe Body, tighten the four screws then recalibrate. * The EC electrode may need recalibrating, so recalibrate it. * Dirty EC electrode: Clean the EC electrode then recalibrate.
  • S
    Steven HernandezAug 27, 2025
    What to do if AquaLink software can't find EIJKELKAMP Measuring Instruments?
    • W
      wmonroeAug 28, 2025
      If the AquaLink software cannot find the Aquameter, it could be due to: * Improperly installed USB drivers: Reinstall the USB drivers carefully following the instructions. * A problem with the USB socket on the PC: Try an alternative socket.
  • A
    Ashley JonesAug 30, 2025
    How to fix GPS on EIJKELKAMP Measuring Instruments?
    • E
      eharrisonAug 30, 2025
      If the GPS Aquameter does not show a position fix, it probably does not have a good enough view of the available satellites. Ensure there are no obstructions between the Aquameter and the open sky. Remember, GPS does not work indoors.
  • S
    Sandra MillerSep 1, 2025
    What does ERROR 02 mean on EIJKELKAMP Aquameter?
    • T
      Tim MorganSep 1, 2025
      ERROR 02 on the Aquameter screen (AP-900 & AP-1000 only) indicates that the Optical DO electrode needs calibrating or the cap needs replacing. Perform a full DO calibration, first at DO Zero then at 100% DO. If that does not cure the problem, replace the Optical DO Cap.
  • L
    Lindsay ToddSep 3, 2025
    What to do if EIJKELKAMP Measuring Instruments can't find Aquaprobe?
    • J
      Justin ChambersSep 3, 2025
      If the Aquameter cannot find the Aquaprobe, there is probably a poor connection. Switch the Aquameter off, disconnect the Aquaprobe plug, ensure there is no debris or moisture in the plug, then re-connect the plug ensuring it is fully inserted and that the screw collar is fully tightened.
  • H
    hillmariaSep 5, 2025
    Why doesn't 'USB CONNECTED' appear on EIJKELKAMP Aquameter?
    • L
      ljohnsonSep 5, 2025
      If the ‘USB CONNECTED’ message does not appear on the Aquameter when connected to a PC, it could be due to: * Dead or incorrectly fitted batteries: Check you have fresh batteries fitted and that they are inserted the correct way round. * A problem with the USB socket on the PC: Try an alternative socket.
  • S
    Shawn GarciaSep 7, 2025
    What does ERROR 01 mean on EIJKELKAMP Aquameter?
    • P
      Phillip KellySep 7, 2025
      ERROR 01 on the Aquameter screen indicates that the pH electrode has dropped below 85% efficiency. Try cleaning the pH electrode and re-calibrating. If that does not cure the problem, replace the electrode.
  • I
    imooreSep 10, 2025
    What does ERROR 07 mean on EIJKELKAMP Aquameter?
    • M
      Mark BrownSep 10, 2025
      ERROR 07 on the Aquameter screen indicates that the Aquaprobe has stopped responding to requests for data from the Aquameter. Check the Aquaprobe plug is fully inserted. Cycle the power to reset the Aquaprobe.
  • S
    Sara AndersonSep 10, 2025
    Why are pH readings inaccurate on EIJKELKAMP Measuring Instruments?
    • B
      Bailey HoodSep 12, 2025
      Slow, inaccurate, or unstable pH and ORP readings, or the inability to calibrate your EIJKELKAMP Measuring Instruments can occur for several reasons: * The electrodes may need recalibrating, so recalibrate them. * The electrodes may need cleaning. * The electrodes may have been allowed to dry out. * The electrodes may be damaged and need replacing. * Water may have entered the electrode socket because the electrode is loose. Remove the electrode, blow out the socket with compressed air, and leave the probe and electrode in a warm place for at least 48 hours to dry out.

Summary

3. BATTERY INSTALLATION AND CARE

4. OVERVIEW OF THE OPERATING SYSTEM

5. CONNECTING AN AQUAPROBE

6. TAKING MEASUREMENTS

6.1. WHAT DOES IT ALL MEAN?

Explains the various parameter readings displayed by the meter.

7. DEPTH MEASUREMENT (AP-1000 ONLY)

8. MEMORY MODE

9. SETTING UNITS OF MEASUREMENT

10. RAPIDCAL CALIBRATION METHOD

10.2. USING RAPIDCAL

Step-by-step guide to performing RapidCal calibration.

10.3. CALIBRATION ERROR CODES

Lists and explains common calibration errors and solutions.

12. GENERAL PROBE MAINTENANCE

13. PH;ORP ELECTRODE CALIBRATION AND MAINTENANCE

13.4. CALIBRATING PH

Detailed procedure for calibrating the pH electrode.

14. DO;EC ELECTRODE CALIBRATION AND MAINTENANCE

14.5. CALIBRATING THE DO ZERO POINT

Step-by-step guide for DO zero point calibration.

14.9. CALIBRATING EC

Detailed procedure for calibrating Electrical Conductivity.

15. TURBIDITY ELECTRODES CALIBRATION AND MAINTENANCE

15.4. TURBIDITY ZERO POINT CALIBRATION

Step-by-step guide for zero point turbidity calibration.

15.6. CALIBRATING THE TURBIDITY 1000 NTU POINT

Procedure for calibrating at the 1000 NTU point.

16. AQUALINK PC SOFTWARE

16.4. UPLOADING DATA FROM YOUR AQUAMETER

Procedure to transfer logged data from meter to PC.

17. SPECIFICATION

20. TROUBLESHOOTING

EIJKELKAMP Aquameter Series Specifications

General IconGeneral
BrandEIJKELKAMP
ModelAquameter Series
CategoryMeasuring Instruments
LanguageEnglish