Guidance and manufacturer’s declaration - electromagnetic immunity
The FlexScan L760T-C is intended for use in the electromagnetic environment specified below.
The customer or the user of the FlexScan L760T-C should assure that it is used in such an environment.
Immunity test IEC60601 test level Compliance level Electromagnetic environment -
guidance
Electrostatic
discharge (ESD)
EN61000-4-2
±6kV contact
±8kV air
±6kV contact
±8kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30%.
Electrical fast
transient / burst
EN61000-4-4
±2kV
for power supply lines
±1kV
for input/output lines
±2kV
for power supply lines
±1kV
for input/output lines
Mains power quality should be
that of a typical commercial or
hospital environment.
Surge
EN61000-4-5
±1kV line(s) to line(s)
±2kV line(s) to earth
±1kV line(s) to line(s)
±2kV line(s) to earth
Mains power quality should be
that of a typical commercial or
hospital environment.
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines
EN61000-4-11
<5% U
T
(>95% dip in
U
T
) for 0.5 cycle
40% U
T
(60% dip in
U
T
) for 5 cycles
70% U
T
(30% dip in
U
T
) for 25 cycles
<5% U
T
(>95% dip in
U
T
) for 5sec
<5% U
T
(>95% dip in
U
T
) for 0.5 cycle
40% U
T
(60% dip in
U
T
) for 5 cycles
70% U
T
(30% dip in
U
T
) for 25 cycles
<5% U
T
(>95% dip in
U
T
) for 5sec
Mains power quality should be
that of a typical commercial or
hospital environment. If the user
of the FlexScan L760T-C requires
continued operation during
power mains interruptions, it is
recommended that the FlexScan
L760T-C be powered from an
uninterruptible power supply or a
battery.
Power frequency
(50/60Hz)
magnetic field
EN61000-4-8
3A/m 3A/m Power frequency magnetic fields
should be at levels characteristic
of a typical location in a
typical commercial or hospital
environment.
NOTE
U
T
is the a.c. mains voltage prior to application of the test level.