Sampling
User Manual Biosen C_line Clinic / GP+
21
Remove excess blood at the end of
the capillary. Use caution so as not
to empty the capillary.
Figure 15 - Removing blood
Place the capillary into the prefilled
micro test tube, close tight, and
shake about 10 times. The capillary
must be completely emptied and
the hemolyzing solution must be
homogenously stained.
Figure 16 - Prepared sample
The sample prepared in this way can be measured immediately.
The glucose and lactate concentration in hemolyzed whole blood, in
closed and not punctured micro test tubes, is stable at room temperature
for a storage period of 5 days. However, it is recommended to store the
samples in a refrigerator if possible and to measure as soon as possible.
6.2 Other sample materials
The glucose and lactate concentration can also be determined in venous
or arterial samples as well as in plasma or serum. The blood collection
systems should be prepared with suitable additives to avoid alteration of
glucose or lactate values over the time period until measurement. If the
blood vessel does not contain a glycolysis inhibitor, the sample must be
processed immediately. The blood may contain the following
anticoagulants / glycolysis inhibitors: heparin, fluoride, EDTA.
Risk of infection, wear suitable gloves!
Take out the sample tube from the fridge and allow it to warm up to
room temperature. Mix the sample carefully by repeated rotating and
rolling.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the EKF Diagnostics Biosen C-Line GP+ and is the answer not in the manual?

EKF Diagnostics Biosen C-Line GP+ Specifications

General IconGeneral
BrandEKF Diagnostics
ModelBiosen C-Line GP+
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals