Important Information
User Manual Biosen C_line Clinic / GP+
1. Important Information
1.1 Safety notes
The following symbols are used in this user manual:
This symbol warns of conditions or handling, which may in certain
circumstances lead to serious injury or even death.
This symbol warns of conditions or handling, which may in certain
circumstances lead to health impairments.
This symbol warns of conditions or handling, which may in certain
circumstances lead to damage to the device or consumables.
It is essential that you read the following notes in order to avoid risks to
persons and damage to the device and other equipment. The handling
of the device requires full understanding and strict observance of this
manual. EKF-diagnostic GmbH does not accept any responsibility for
damage arising from non-observance of the following notes.
The device processes samples of human origin. This
involves potential risk of infection. Particularly the sample cannula, the
exchanger, the chip sensors, the tray and the waste container can be
contaminated. Gloves should always be worn when operating the
device.
Danger of fatal electric shock!
Under no circumstances should you open the device. There are no
components inside which require servicing or maintenance. Repairs
must only be performed by authorized service staff.
Never use a mechanically damaged device.
An isolated ground receptacle is mandatory for operating the device.
Check whether the mains voltage and frequency printed on the type
label match your mains socket.
The device is not insulated against fluid ingress. Note
the maintenance instructions in section 12, page 44. If a fluid spillage is
suspected pull the mains plug immediately.