EasyManua.ls Logo

elexxion claros pico - Technical data base unit

elexxion claros pico
34 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
claros pico user manual V 2.1 page 27 of 34
12.5
Technical data base unit:
Power supply: 5V/DC, < 3.0 A
(with Friwo FW7556M/05 mains adapter supplied)
Fuses: No accessible fuses
Operating mode: Intermittent operation / 2 min ON / 1 min OFF
Protection class: I application part type B / IP 00
Wavelength: 808 nm +/- 10 nm
Max power output (pulse): 5 W
Max. power output (CW): 1.0 W
Pulse frequency: CW – 20.000 Hz
Pulse duration: 26 µs,50 µs, 150 µs
Aiming beam: 650 nm +/- 5 nm / < 1mW
Operating conditions: Temp. 15°C – 35°C
Rel. humidity 20 – 85 %
Air pressure: 800 – 1100 hPa
Storage/transport conditions: Temp. 5°C – 50 °C
Rel. humidity 10 – 85 %
Air pressure: 800 – 1100 hPa
12.6
Accuracy of values displayed:
output: +/- 20 % from value specified