Table1:GuidanceandManufacturer’sDeclaration
ElectromagneticEmissions
ELITONEisintendedforuseintheelectromagneticenvironmentspecifiedbelow.Thecustomerortheuser
shouldassurethatitisusedinsuchanenvironment.
EmissionsTest Compliance ElectromagneticEnvironment‐ Guidance
RFemissionsCISPR11 Group1
ELITONE usesRFenergyonlyforitsinternalfunction.Therefore,
itsRFemissionsareverylowandarenotlikelytocauseany
interferenceinnearbyelectronicequipment.
RFemissionsCISPR11 ClassB
ELITONE issuitableforuseinallestablishmentsincluding
domesticandthosedirectlyconnectedtothepublicsupply
networkthatsuppliesbuildingsusedfordomesticpurposes.
Table2:GuidanceandManufacturer’sDeclaration
ElectromagneticImmunity
forallequipment
ELITONEisintendedforuseintheelectromagneticenvironmentspecifiedbelow.Thecustomerortheuser
shouldassurethatitisusedinsuchanenvironment.
ImmunityTest TestLevel Compliance ElectromagneticEnvironment‐Guidance
ElectrostaticDischarge
(ESD)IEC61000‐4‐2
±8kVContact
±15kVAir
±8kVContact
±15kVAir
Floorsshouldbewood,concreteorceramic
tile.Iffloorsarecoveredwithsynthetic
material,therelativehumidityshouldbeat
least30%.