AM852 Veterinary Anaesthesia machine User Manual
27
8.Trouble shooting
Warning: Please make eligible service representative repair the system.
Fault symptom and
alarm info
Possible cause
Troubleshooting
The animal respiratory
circuit leak gas
APL valve disclose by manual
control
Close APL valve
Absorber fix untightly
Fix it again
Corrugated hose mangle or joint
loose
Replace hose or fix again
Inhale or exhale unidirectional
valve cover doesnt screw down
Screw again
Manual switch doesnt work
Please contact with our company
Airway pressure is far
high when manual
respiration
APL valve adjust isnt accurate
Adjust APL valve to right
position