AM852 Veterinary Anaesthesia machine User Manual
30
10. Technical Specification
Remark: All technical specifications are rating, may be changed at any time, and do not message
again.
Gas Supply
pipeline gas
O2N2O
the output pressure of
regulator
280kPa-600kPa
security valve discharge
pressure
600 kPa
the pressure of pipeline
inlet
280kPa-600kPa
Note Gas providing must be medicinal level.
Note When the medical gas pipeline supply pressure is lower than 0.15Mpa,the
anaesthetic system will cease to deliver gas.
Flowmeter
Gas
scale
O2
0.0L/min 10L/min
N2O
0.0L/min
10L/min
Precision: At 20 , atmospheric pressure is101.3kPa,the precision of flowmeter is level
2.5.
O2 flushing 25L/min 75L/min
O2 supply trouble alarm O2 pressure lower than 0.14MPa
N2O cut off O2 lower than 0.1MPa