EasyManua.ls Logo

ELM AM852 - Page 31

Default Icon
35 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
AM852 Veterinary Anaesthesia machine User Manual
32
5. Unscrew the APL valve until it is fully open.
6. Set the flow to 30l/m.Check that the pressure reading is between 1 and 3cm H
2
O.
7. Set the flow to 30l/m (from a separate source if necessary) and check that the pressure
reading is between 1 and 5cm H
2
O.
8. For semi-closed circuits, check the continuity of the inner tube by passing a flow of
20l/m along the inner tube with the APL valve closed. Block the outlet for a time
sufficient to allow the bag to partially fill. The reservoir bag shoule then collapse by
Venturi action when the animal outlet is opened to the atmosphere. If the bag
continues to fill, there is a break in the inner tub
دﯾﯾﺎﻣرﻓ ﮫﻌﺟارﻣ.WWW.ELM-FAN.IRتﯾﺎﺳ ﮫﺑ رﺗﺷﯾﺑ دراوﻣ دوﻠﻧاد یارﺑ
To download more items, go to ELM-FAN.IR