AM852 Veterinary Anaesthesia machine User’ Manual
6
and the CO
2
check and test (according to EN864 or ISO9918) in this device.
User should equip CO
2
check and test according with EN740 standard.
Importance : According to the stipulation of European standards EN740 and international
standards IEC 60601-2-13/ISO8835-1, it is essential to check and test the
anesthetic (according to ISO11196) when using Anaesthesia evaporator. User
should equip Anaesthesia evaporator monitor according with EN740 standard.
Warning: Please read this manual in detail and master the content before operating the
product.
Warning
:
Ensure the avirulence of air feed tube, bag and breathing loop assembly, and the
followings must not emerge:
·Cause allergic reactions in the animal’s body.
·React with Anaesthesia gas or anesthetic to produce dangerous byproducts.
Warning:The danger of explosion. Ensure not to use flammable anesthetic in AM852 system.
Warning
:
Do not use anlistatig breathing tube or face mask. It is possible to cause burning
when using this species of breathing tube or face mask near surgery equipments of
high frequency.
Warning
:
In order to protect the animal, the following must be ensured.
·Ensure the operations of all life supports and monitor assembly are proper.
·Ensure manual ventilation equipment convenient .
Warning : If the anaesthetic system is to be equipped with an anaesthetic ventilator, the
anaesthetic ventilator shall comply with the requirements of ISO 8835-5.
Warning: a malfunction of central gas supply system may cause more than one or even all
devices connected to it to stop their operation simultaneously.