F_Get_Targets_Vcc
F_Get_Targets_Vcc - Get Vcc in [mV] supplied target device.
VALID FPA index - ( 1 to 64 )
Syntax:
MSPPRG_API INT_X F_Get_Targets_Vcc( void );
Return value:
INT_X - Vcc in milivolts e.g 3000 -> 3.0 V
or (-1) if USB-FPA is not active.
-2 (0xFFFFFFFE) - FPA_INVALID_NO
F_Set_fpa_io_state
F_Set_fpa_io_state - Set state of the Reset, Vcc and Jtagl lines
VALID FPA index - ( 1 to 64 ) or 0 (ALL FPAs) executed sequentially.
Syntax:
MSPPRG_API INT_X F_Set_fpa_io_state( BYTE jtag, BYTE reset, BYTE Vccon void );
jtag -> TMS, TCK, TDI (output from FPA)
0 -> DEFAULT_JTAG_3ST
1 -> DEFAULT_JTAG_HI
2 -> DEFAULT_JTAG_LO
reset -> 0 -> output from FPA RESET - LO
1 -> output from FPA RESET - HI
VccOn -> 0 -> output Vcc from FPA - OFF
1 -> output Vcc from FPA - ON
(level 2.2V to 3.6V set in CFG_VCCINDEX )
74