Parameter FREE CHLORINE (ORGANIC AND INORGANIC) FOR SALT WATER
Measuring range 0-10 mg/l (0-10 ppm) resolution: ± 0.05
Connection 2 wires (+red; -black)
Measuring system amperometric - 2 electrodes (platinum/silver)
Ph working range 6-8 pH
Run-in-time
First polarization: 2 h about
Next polarizations: 50 min. about
Response time T
90
: 2 min. about
Zero point
adjustment
See Operating manual: “Calibration”
Slope calibration See Operating manual: “Calibration”- DPD1 method
Alcalinity 100 ppm
Working
temperature
5-40° C (41-104°F)
Pressure 0.4 - 5 bar (5.8 - 72.5 PSI)
Cable (standard) 2 m (6.6 ft); 1 m if assembled on panel
Working flow 40 l/h
Suitable as probe
holder for
pH, ORP and temperature temperature
Fittings for
connection to the
sample pipeline
6x8
Material
Electrode: platinum/silver
Measurement cell: metacrylate (PMMA)
Mounting On flat vertical surface (panel, support, etc.).
Storage Frost and dry protected (5-40° C)
Maintenance
Regular control of the signal
SHORTEN THE MAINTENANCE INTERVALS APPROPRIATELY DEPENDING
ON WATER QUALITY.