Installation
Upon receipt, check for any visible damage which may have occurred in
transit. If any signs of damage are found, then retain all packing material
and inform the carrier and the Company or its agent from whom the unit
was purchased within two working days.
The Primo Interferential 960/Multidyne 970 must only be used with an EMS
SLA9000 power supply (as supplied with the unit) which is connected to a
mains supply of 100-240V ac. A power cord appropriately rated/approved
for the country of use must be used.
The SLA9000 power supply must only be connected to a mains supply with
a protective earth conductor. If the integrity of the earth connection is in
doubt, do not connect it to the mains supply (risk of electric shock with type
B applied parts). The unit must not be positioned in such a way that the
mains plug cannot easily be unplugged – the mains plug is the main
disconnect device.
The Primo Interferential 960/Multidyne 970 unit is supplied with four
medium-sized electrotherapy electrodes with their associated patient lead.
Plug the patient lead into the socket on the front of the unit and connect the
electrodes to the yellow and blue cables.
Operation of the unit in close proximity (less than 1 metre) to shortwave
therapy equipment or radio-frequency mobile communication equipment
could result in the output of the Interferential 960/Multidyne 970 being
affected.
Permissible environmental conditions of use:
Temperature 10 to +35°C
Relative humidity 10 to 80%
Atmospheric pressure 500 to 1060hPa
Permissible environmental conditions of transport and storage:
Temperature -10 to +35°C
Relative humidity 5 to 95%
Atmospheric pressure 500 to 1060hPa