Appendix B – Technical specification
General
Power input (SLA9000) 100-240V ac 1.5A 50-60Hz
(EMS960/970) 18V, 3.33A (from external PSU SLA9000)
Classification (EN60601-1) Class 1, Type BF
Fuse Internal T3.15A
Size (height x width x depth) 108 x 237 x 333 mm
Weight 1.3 kg
Treatment programs 16 user-defined set-ups.
Constant Current (CC) measured into 500 Ohm, Constant Voltage (CV)
into open circuit. Operation outside a safe range is prevented by the
electrode fault monitor alarm.
Interferential 4-pole (960 & 970)
Carrier frequency 2 kHz, 4 kHz or 8 kHz
AMF 0 – 250 Hz in 1 Hz increments
Swing pattern 1|1, 6|6 or 6^6
Vector 10s, 20% both channels
Output type CC 0-100mA peak
CV 0-70V peak
Interferential 2-pole (960 & 970)
Carrier frequency 2 kHz, 4 kHz or 8 kHz
AMF 0 – 250 Hz in 1 Hz increments
Swing pattern 1|1, 6|6 or 6^6
Output type CC 0-100mA peak
CV 0-70V peak
Russian stimulation (970 only)
Carrier frequency 2.5 kHz
Modulation frequency 1 – 100 Hz
Surges 1:1 to 1:5
Output type CC 0-100mA peak
CV 0-70V peak