6
AUTOMATIC MANUAL
The following automated process
1
can be used to
achieve an A0 Level of 3000:
2 min
Pre-clean with cold tap water
2
.
Drain
5 min
Clean at 55°C with tap water
2
and 0.5%
cleaning solution.
Drain
3 min
Rinse and neutralize with cold
demineralized water
3
.
Drain
2 min
Rinse with cold demineralized water
3
.
Drain
3 min
    
with deionized water at 93°C minimum.
Drain
20 min
(at least)
Dry at 100°C.
DO NOT use any ultrasound bath cleaning
procedure with the handpieces: it may destroy the
products.
The following process is intended ONLY for EMS
PIEZON
®
Instruments
15 s
For product with lumens, flush all lumens
with a spray gun (water jet gun, with a static
water pressure of no more than 2 bar) with
cold tap water.
10 min
Place products in a metal sieve and sonicate
in 0.5% cleaning solution of neodisher
4
in
deionized water at 40°C
5
. Make certain that
  
(use a syringe if needed).
15 s
Rinse all lumens by flushing with a spray
gun (water jet gun, with a static water
pressure of no more than 2 bar) with cold
deionized water.
10 s
Rinse the whole product under cold running
deionized water.
DISINFECT
ASP CIDEX
®
OPA solution must be used undiluted,
within its useful and shelf life, and in compliance with
the manufacturer’s warnings and instructions for use.
5 min
Immerse the product completely into CIDEX
®
OPA solution at minimum 20°C. Make certain
      
solution (use a syringe if needed).
1 min
(each rinse)
The ASP CIDEX
®
OPA disinfectant requires
a total of three rinses : keep the product
totally immersed, and use a large volume
6
of fresh water.
Do not reuse the water for rinsing or any other purpose.
Residues of disinfectant may cause serious side effects.
DRY
Use an air pistol (compressed air) to fully dry the lumen
and the whole product until no more residues of water
are present (visible or detectable).
INSPECT
If stains are still visible on the product after cleaning/disinfection, the entire cleaning/disinfection procedure
must be repeated. Products with visible damage, chip/flake loss, corrosion or deformation must be disposed of (no
further use is allowed). Check also the integrity of O-rings and gaskets and replace if damaged or deformed.
Verify the product to be fully dry. In case of detection of residues of water, remove these using an air pistol (clean
compressed air). Fully dry the lumen and the whole product until no more residues of water are present (visible or
detectable).
1
For example: Miele Professional G 7836 CD with Miele Rack E429
2
cold tap water =
16°C +/- 2°C
3
cold demineralized water =
20°C +/- 2°C
4
For example: MediClean Dental,
Dr. Weigert Hamburg
5
For example: Bandelin, Sonorex 1028K, 35kHz
6
e.g. 2 gallons

Other manuals for EMS AIRFLOW EL-185

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the EMS AIRFLOW EL-185 and is the answer not in the manual?

EMS AIRFLOW EL-185 Specifications

General IconGeneral
BrandEMS
ModelAIRFLOW EL-185
CategoryDental equipment
LanguageEnglish

Related product manuals