PROCESS
PREPARE
20 s - Immediately after use, rinse the lumen(s) line of the handpiece/instrument with water.
Coarse soiling must be removed immediately after application.
Safely transport to the reprocessing area to avoid any damage to the products, contamination of the environment
or of the persons involved in the reprocessing process.
Cleaning is to be performed within 1 hour from use.
Wear personal protective equipment according to the type of preparation.
PIEZON
®
Handpieces:
Remove any installed instrument.
Remove the nose cap and separate the light guide and O-ring.
AIRFLOW
®
and PERIOFLOW
®
Handpieces:
Always carry out handpiece powder unclogging and
check for both lumens (water and powder) clearance
before proceeding. Use Easy Clean.
PIEZON
®
Instruments with CombiTorque
®
:
Separate the PIEZON
®
Instrument and the
CombiTorque
®
but keep each original instrument
and Combitorque
®
together throughout their life.
PIEZON
®
Endochuck:

Remove the screw and separate the small O-ring.
Any product can be cleaned manually or automatically by washer or disinfector.
EMS recommends the use of an ISO 15883 compliant automatic washer-disinfector (WD) for optimal
effectiveness and service life.
AUTOMATIC MANUAL
CLEAN, DISINFECT AND DRY
The washer-disinfector must comply with ISO 15883,
must have suitable baskets to hold small fragile products
and must have rinsing connections with diameter of
approximately 16mm for the attachment to product
lumen.
Correctly place the product into a suitable rack, connect all
lumens to the rinsing connectors and start the automated
cleaning.
Instructions of the manufacturer of the washer
disinfector must also be observed.
In case of use of a chemical disinfectant, follow
carefully the instructions provided by the disinfection
solution manufacturer.
CLEAN
The following process can be used with any EMS
product:
Brush the devices in a cleaning solution of 0.5% neodisher
1
in deionized water at 40°C with a suitable soft bristled
brush
2
until all visible residue has been removed.
15 s
For products with lumens, flush all lumens
with a spray gun (water jet gun, with a static
water pressure of 2 bar) with cold tap water.
15 min
Place the product in 0.5% solution of
neodisher
1
with deionized water at 40°C.
 
cleaning solution (use a syringe if needed).
15 s
Rinse all lumens by flushing with a spray gun
(water jet gun, with a static water pressure
of 2 bar) with cold deionized water.
10 s
Rinse the whole product under cold running
deionized water.
01 PREPARE 02 CLEAN 03 DISINFECT 04 DRY 05 INSPECT 06 PACK 07 STERILIZE 08 STORE
1
For example: MediClean Dental,
Dr. Weigert Hamburg
2
Medisafe MED100.33
EN

Other manuals for EMS AIRFLOW EL-185

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the EMS AIRFLOW EL-185 and is the answer not in the manual?

EMS AIRFLOW EL-185 Specifications

General IconGeneral
BrandEMS
ModelAIRFLOW EL-185
CategoryDental equipment
LanguageEnglish

Related product manuals