16
MODE Interpretations
Modulated
Intensity and
Pulse Width
(SD)
so that the intensity is always increasing while the pulse width is decreasing and
equal to setting). Total cycle time is 6 seconds. Pulse rate (from 2 to 150Hz) and
pulse width (from 50 to 300µs) are fully adjustable.
Bi-Pulse
(Bi-Pulse)
rate of Channel 1 is fixed at 4Hz) while delivering 100 pulses per second to Chan-
nel 2 (i.e. the pulse rate of Channel 2 is fixed at 100Hz). Stimulation is burst on
Pulse width (from 50 to 300 µs) is fully adjustable.