EasyManua.ls Logo

Endress+Hauser Liquipoint FTW 31 - FEW 58 NAMUR Output Implementation

Endress+Hauser Liquipoint FTW 31
28 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Liquipoint T FTW 31, FTW 32
Endress + Hauser 11
Compact instrumentation with FEW 58
To be used with isolating amplifiers acc. to NAMUR (IEC 60947-5-6)
Output signal jump from high to low current on limit (H-L-edge).
Signal transmission on a two-wire line:
H-L-edge 2.2 ... 6.5 mA / 0.4 ... 1.0 mA
When using a multiplex the cycle time must be set to a minimum of 2 s.
L00-FTW3xxxx-04-05-xx-en-005
Connecting the FEW 58 electronic insert.
Power supply (FEW 58)
Refere to "Technical Data" date sheet of the connected isolating amplifier acc. to
NAMUR (IEC 60947-5-6)
Signal on alarm
Output signal with damaged sensor: < 1.0 mA
1
+
2
FEL 58
E
X
E
X
I
IEC 60947-5-6 (NAMUR)IEC 60947-5-6 (NAMUR)