OPTIONAL POLI – X ACCESSORIES
P-MF Maintenance Free Prophy Angle
P-106 Screw On Rubber Cups 144 / pkt
CARE FOR YOUR MAINTENANCE-FREE ANGLE STERILIZATION PROCEDURES:
AFTER EACH PROPHY:
1. Discard used rubber cup.
2. Rinse abrasive paste from head area with water.
3. Thoroughly clean the outside of angle with disinfectant.
4. Autoclave angle - not more than 275 ºF (135 ºC) over 20 minutes.
5. After sterilization cycle is complete, install a new Disposable Rubber Cup and attach angle to
handpiece. You are now ready for your next prophy.
CAUTIONS AND WARNINGS:
1. Do not attempt to disassemble.
2. DO NOT SUBMERGE IN LIQUIDS, INCLUDING ULTRASONIC SOLUTIONS.
3. Do not heat over 275 ºF (135 ºC).
4. Use only Engler Care Free Prophy Rubber Cups. Other brands will not properly seal the angle,
causing premature wear and voiding the warranty.
5. Replace as necessary.
YOUR CARE-FREE ANGLE IS WARRANTED AGAINST DEFECTS IN MATERIAL AND
WORKMANSHIP FOR 6 MONTHS. A COPY OF OUR INVOICE OR PICKING TICKET WILL
BE REQUIRED AS PROOF OF PURCHASE.
WARRANTY IS VOID IF:
1. Engler Care-Free rubber cups are not used exclusively.
2. Sterilization procedure is not followed properly.
3. The angle has been submerged in any liquid.
4. The angle has been damaged or abused.
5. Damaged due to use at high speed.