ADJUSTMENT OF THE HORIZONTAL POSITION ON
THE MANUMED OSTEO
The Manumed Osteo with electric section
adjustment are fitted with programmable
operating buttons.
A. Use the electric operation A and B to adjust the
sections up and down independently.
The positions of the back-section and the
sitting-section are programmable. Press
buttons 3 and 4 (both buttons from B) in order
to recall the programmed position
B. Reprogramming the positions of the back-
section and the sitting-section
a. Adjust the sections to the desired position.
b. Press keys 1 and 2 simultaneously
c. Confirm the settings by pressing key 3 and
4 within 2 seconds. The new settings are
acknowledged with a beep.
A = electric operation back-section
B = electric operation sitting-section
1 = back-section up
2 = back-section down
3 = sitting-section up
4 = sitting-section down
ADJUSTMENT OF THE ARM COUCH SECTIONS
The arm sections can be adjusted for height by
loosening the knob and sliding the arm sections
along the runners.