Recommended separation distances between portable and mobile RF communications equipment
and the 6-series
The 6-series is intended for use in an electromagnetic environment in which radiated RF disturbances are
controlled. The customer or the user of the 6-series can help prevent electromagnetic interference by
maintaining a minimum distance between portable and mobile RF communications equipment
(transmitters) and the 6-series as recommended below, according to the maximum output power of the
communications equipment.
Rated maximum
output power
of transmitter
[ W ]
Separation distance according to frequency of transmitter
[ m ]
150 kHz to 80 MHz
d = [3,5/V
1
]√P
80 MHz to 800
MHz
d = [3,5/E
1
]√P
800 MHz to 2,5 GHz
d = [7/E
1
]√P
0,01
0,12
0,12
0,23
0,1
0,37
0,37
0,74
1
1,17
1,17
2,33
10
3,69
3,69
7,38
100
11,67
11,67
23,33
For transmitters rated at a maximum output power not listed above, the recommended separation
distance d in meters (m) can be estimated using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people
14.7 Technical Data
Mains voltage : 100 240 V ± 10%
Mains frequency : 50 / 60 Hz
Main unit:
Dimensions : 39 x 37 x 12 cm (w x d x h)
Weight : 4.75 Kg
Vacotron reservoir dimensions : 39 x 37 x 6 cm (w x d x h)
Weight : 1.5 Kg
Environmental conditions for transport and storage:
Temperature : -10° to +50° C
Relative humidity : 10 to 90 % (contained in original packaging)
Atmospheric pressure : 500 to 1060 hPa
Environmental conditions for normal use:
Temperature : 10° to 40° C
Relative humidity : 10 to 90 % non condensing
Atmospheric pressure : 500 to 1060 hPa
Technical modifications reserved

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Enraf Nonius Sonopuls 692 and is the answer not in the manual?

Enraf Nonius Sonopuls 692 Specifications

General IconGeneral
BrandEnraf Nonius
ModelSonopuls 692
CategoryMedical Equipment
LanguageEnglish

Related product manuals