EasyManuals Logo

Enraf Nonius Sonopuls 692 User Manual

Enraf Nonius Sonopuls 692
106 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #98 background imageLoading...
Page #98 background image
96
1600751-4D
13.13 Vacuum pulsed mode
p Vacuum pressure
r Range
t
h
Hold time
13.14 Ultrasound
Ultrasound parameters:
f
p
Pulse frequency, 100 Hz fixed
t Pulse duration, set by duty cycle
5 80 % and 100 % (100% is
continuous wave)
t 0.5 8 ms
RTPA 20 1.25
14 Specifications
14.1 Ultrasound parameters
See fig 13.14 for details.
Generator
Peak output Amplitude:
Duty cycle 5 50 % : 0 3 W/cm²
Duty cycle 80 % : 0 2.5 W/cm²
Duty cycle 100 % : 0 2 W/cm² (continuous wave)
Peak output power for 5 cm² applicator:
Duty cycle 5 50 % : 0 15 W
Duty cycle 80 % : 0 12 W
Duty cycle 100 % : 0 10 W (continuous wave)
Peak output power for 0.8 cm² applicator:
Duty cycle 5 50 % : 0 2.4 W
Duty cycle 80 % : 0 2 W
Duty cycle 100 % : 0 1.6 W (continuous wave)
Output meter uncertainty : ± 20 % for any output above 10 % of maximum
Pulse frequency : 100 Hz ± 1 % fixed
Duty cycle : 5 80 % and 100 % (100 % is continuous wave)
Pulse duration : 0.5 8 ms ± 10 % (set by duty cycle)
Temporal Peak to Average Ratio (RTPA):20 1.25 ± 10 % (set by duty cycle)
Treatment timer : 0 - 30 min ± 0.1 min, linked to contact control
Contact control level : 65%
5 cm² Applicator
Ultrasound frequency:
1 MHz : 0.98 MHz ± 2 %
3 MHz : 3.1 MHz ± 2 %

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Enraf Nonius Sonopuls 692 and is the answer not in the manual?

Enraf Nonius Sonopuls 692 Specifications

General IconGeneral
BrandEnraf Nonius
ModelSonopuls 692
CategoryMedical Equipment
LanguageEnglish

Related product manuals