What to do if my Esco Airstream PCR-3A cabinet does not start?
- VvsuarezNov 9, 2025
If your Esco Laboratory Equipment cabinet isn't starting (no LCD, button, fan, light, or UV function), first, check if there's power at the wall socket using a voltmeter or test pen. Then, ensure the power cord is properly connected to both the wall and the unit. If there's still no power, and if you have a PCR model, check the circuit breaker located beside the power inlet. If the breaker tripped, do not reset it before inspecting all electrical components and wiring. Also, check the connection to the main power supply inside the electrical panel, ensuring the voltage is within the correct range (230 VAC or 115 VAC ± 10%). For PCR models, a defective power supply or main board could also be the cause. Disconnect the 5-pin connector to check the SMPS output. If the voltage is out of range,...