Thank you for purchasing this Esco Polymerase Chain Reaction
Cabinet. Please read this manual thoroughly to familiarize yourself
with the many unique features and exciting innovations we have built
into your new equipment. Esco provides many other resources at our
website, www.escoglobal.com, to complement this manual and help
you enjoy many years of productive and safe use of your Esco
products.
Esco Technologies, Inc.
903 Sheehy Drive, Suite F, Horsham, PA 19044, USA
Tel 215-441-9661 • Fax 484-698-7757
escolifesciences.us • eti.admin@escoglobal.com
Rest of World
Esco Micro Pte. Ltd.
21 Changi South Street 1 • Singapore 486 777
Tel +65 6542 0833 • Fax +65 6542 6920
www.escoglobal.com mail@escoglobal.com
Esco Polymerase Chain Reaction Cabinets
(SCR-2A, PCR-3A, PCR-4A)
Service Manual Version E - Released January 2021
North America
Service Manual
Polymerase Chain Reaction
Cabinets
DOMINIQUE DUTSCHER SAS

Questions and Answers

  • J
    Joseph GreeneNov 9, 2025
    What to do if my Esco Airstream PCR-3A cabinet does not start?
    • V
      vsuarezNov 9, 2025
      If your Esco Laboratory Equipment cabinet isn't starting (no LCD, button, fan, light, or UV function), first, check if there's power at the wall socket using a voltmeter or test pen. Then, ensure the power cord is properly connected to both the wall and the unit. If there's still no power, and if you have a PCR model, check the circuit breaker located beside the power inlet. If the breaker tripped, do not reset it before inspecting all electrical components and wiring. Also, check the connection to the main power supply inside the electrical panel, ensuring the voltage is within the correct range (230 VAC or 115 VAC ± 10%). For PCR models, a defective power supply or main board could also be the cause. Disconnect the 5-pin connector to check the SMPS output. If the voltage is out of range,...
  • D
    Debra SandersNov 19, 2025
    Why is the airflow not at nominal velocity in my Esco Airstream PCR-3A Laboratory Equipment?
    • N
      Nathan HutchinsonNov 19, 2025
      If the airflow in your Esco Laboratory Equipment is not at the nominal velocity, first check for external air disturbances such as air conditioning vents, open windows, or people walking nearby. Make sure nothing is blocking the pre-filter. If there are no obstructions, switch on the blower fan and measure the AC voltage at the motor voltage sampling port. If the motor voltage is lower than the nominal value from the factory test report, adjust the speed control to get the specified output voltage. Also, check if the building supply voltage is low, especially with new cabinets, and adjust the speed control accordingly. If the cabinet has been used for over a year, refer to the last motor nominal voltage from the certification report. If the voltage is the same, the filter may be loaded, or...
  • B
    Brenda FerrellDec 1, 2025
    How to troubleshoot blower not functioning in Esco Airstream PCR-3A Laboratory Equipment?
    • T
      Timothy PayneDec 1, 2025
      If the blower in your Esco Laboratory Equipment isn't working, first, ensure the fan is switched on via the control panel (enter the PIN if required for PCR models, default is 0001, or switch the rocker switch for SCR models). If the LED doesn't light up, check the connection from the control panel to the main board or replace the membrane. For SCR models, check for loose wiring at the back of the switch. If the fan still doesn't work on PCR models, check fuse F1 on the relay board and replace it with F5 as a temporary solution if blown. Next, unplug the cabinet, open the electrical panel, and add a jumper wire between the two cables going to the speed control. If the fan now works, replace the speed controller. If not, the capacitor may be faulty; disconnect its cables and measure the cap...

Summary

Introduction

Safety Warning

Essential safety precautions for operating and working around the equipment.

Chapter 1 - Re-Certification by Service Personnel

Unit Re-Certification

Key procedures and checks involved in re-certifying the cabinet.

Replacement of Filter

Detailed instructions for safely replacing the HEPA/ULPA filter.

Replacement of Blower

Guide on how to replace the blower motor and housing.

Chapter 2 - Troubleshooting

Cabinet does not start

Diagnostic steps for issues preventing the cabinet from powering on.

Blower does not function

Troubleshooting guide for when the blower fan fails to operate.

Airflow is not at nominal velocity

Steps to address airflow issues and achieve specified velocity.

Esco Airstream PCR-3A Specifications

General IconGeneral
TypePCR Cabinet
Filter TypeHEPA
UV LampYes
HEPA Filter Efficiency99.99% at 0.3 microns
Noise Level65 dBA
Power Supply220-240 V AC, 50/60 Hz