Installation
When unit is removed from skid, the supplied casters must be installed on the
bottom of unit. Place casters with locking brakes on front corners of unit.
The unit has been thoroughly bench-tested and is shipped to the customer ready to
operate. Move the unit into position and lock Caster wheels. Once the position of
the unit has been established, the electrical and plumbing requirements must be
completed by the user.
The power modules inside the unit have been bolted to the bottom of chassis for
shipping. They can remain bolted. However, if either has to be removed for any
reason, the filter screen door on the back of the unit has to be opened. Once the
filter is removed, the screws and bolts (2’’ in each bracket) in hold-down brackets
can be removed. These need not be replaced for normal use and movement of unit.
Leave hold-down bracket fixed to power modules.
Electrical
The E992 unit is wired for 230 VAC. The rating is 9500 watts, single phase
Each Heated Storage Tank (HST) is rated for 1,650 watts(included above),
120VAC, 50/60HZ and is supplied with a cord with a NEMA 5-15 plug. A 15
amp, 125 volt fuse is located underneath the control box. The power cords are
intended to be plugged into the master electrical panel located on the left side of
the E992 unit. The system is designed to be wired at this panel. An electrical
diagram is located inside the panel door and also as an attachment to this manual.
DO NOT OPERATE UNIT WITHOUT PROPER GROUNDING.
Each heated storage tank (HST) has a separate communication cord with an
amphenol connector to be connected to the sockets at the rear of the washer as
follows: