Guidance and manufacturer’s declaration - electromagnetic immunity
The synapsis wave 2 is intended for use in the electromagnetic environment
specied below. The customer or the user of the synapsis wave 2 should
assure that it is used in such an environment(wave healthcare environment
– excluding outdoors, vehicles and public places).
Immunity test IEC 60601 test level
Compliance
level
Electromagnetic
environment -
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 8 kV contact
± 15 kV air
± 8 kV
± 15 kV
Floors should be
wood, concrete or
ceramic tile. If floors
are covered with
synthetic material,
the relative
humidity should be
at least 30 %.
magnetic eld
Power
frequency
(50/60) Hz
magnetic eld
IEC 61000-4-8
30 A/m 30 A/m
Power frequency
magnetic elds
should be at levels
characteristics of a
typical location in a
typical commercial
or hospital
environment.
Conducted RF
IEC 61000-4-6
V1 = 6 Vrms
150 kHz – 80 MHz
V1 =
6 Vrms
Radiated RF
IEC 61000-4-3
E1 = 10 V/m
for wave healthcare
environment
80 MHz – 2,7 GHz
10 V/m
Field strengths from
xed RF
transmitters, as
determined by an
electromagnetic site
survey, should be
less than the
compliance level in
each frequency
range. Interference
may occur in the
vicinity of
equipment