Calibration:
Acousticalcalibrator(Extech407744or equivalent)
Calibrationcheckfrequencyis1000Hz.
Nominalcalibrationlevelforthefreefield:94.1dB
Nominalcalibrationlevelforthediffusefield:94.0dB
Frequencyforacoustictesting: 8000Hz
Warmuptime: 2min
Samplinginterval: Bargraphindication
125msapprox.
Numericindication
1secapprox.
Datarecordcapacity: Datacanbestoredinmemory.
Max.32000readingscanbestored
Max.255blocks
LCDDisplay
Displayscreens:
4digitnumerical indicationof soundlevel,from30.0to130.0 dBwith0. 1dBresolution.
Bargraphindicationofcurrentsoundlevelwith1dBresolution.
Soundlevelrange
indicator:30–90dB,40–100dB,50–110dB,60–120dBor70–130dBinfive
ranges.
Timedisp lay ;yearmonthdayandhour:minute:second.
Displayupdaterate:1second
Display’sstartupindication:Dependsontheconditionofmeterwhenitwaslastswitchedoff.
Warningindications:
Outofrangeindications:
OVERdisplayed
atupperlimitoftherange
UNDERdisplayedatlowerlimitoftherange
Outputs
ACoutput(usingselectedfrequencyweighting)
Outputvoltage: 2Vrms(atfullscaleoftherange)
Outputimpedance: 5kΩ
Loadimpedance: ≥1MΩ
DCoutput
Outputvoltage: 10mV/dB
Outputimpedance: 5kΩ
Loadimpedance: ≥1MΩ
I/Oconnector: Soundlevelmetercontr olfrom,anddat aoutputto,PC(USB)
Alarmoutput: 5Vdc,typical