General Specifications
Diode Test Test current of 1mA maximum, open circuit voltage 2.8V DC
typical
Continuity Check Audible signal will sound if the resistance is less than
approximately 150Ω
Input Impedance 10M
AC Response Average responding
ACV Bandwidth 50Hz to 1kHz
DCA voltage drop 200mV
Display 3 ½ digit, 2000 count LCD, 0.9” digits
Auto-Power Off 15 minutes (approximately)
Overrange indication “1” is displayed
Polarity Automatic (no indication for positive polarity); Minus (-) sign for
negative polarity.
Measurement Rate 2 times per second, nominal
Low Battery Indication is displayed if battery voltage drops below operating
voltage
Battery One 9 volt (NEDA 1604) battery
Fuses mA, µA ranges; 0.2A/250V fast blow
A range; 20A/250V ceramic fast blow
Operating Temperature 5ºC to 40ºC (41ºF to 104ºF)
Storage Temperature -20ºC to 60ºC (-4ºF to 140ºF)
Operating Humidity Max 80% up to 31ºC (87ºF) decreasing linearly to 50% at
104ºF (40ºC)
Storage Humidity <80%
Operating Altitude 2000 meters (7000ft.) maximum
Weight 342g (0.753lb) (includes holster)
Size 187 x 81 x 50mm (7.36” x 3.2” x 2.0”) (includes holster)
Safety For indoor use and in accordance with the requirements for
double insulation to IEC1010-1 (2001): EN61010-1 (2001)
Overvoltage Category III 600V and Category II 1000V,
Pollution Degree 2.