Each inlet to the rotary valve is equipped with a filter. If it becomes clogged, it
should temporarily be removed. Do not operate continuously without a filter. Any
particulates present in the sample will drastically shorten the valve’s life.
If the actuator loses proper alignment with the valve body, it may resemble a
blocked valve.
Valve will not rotate
May result from an improperly connected valve actuator, either the DC power
cord or the Serial COM cable (three pin push on connector) that commands the actuator.
A faulty valve actuator.
A valve rotor that has seized.
Valve selects the wrong stream
Indicates a problem with valve alignment. Refer to the Multiposition Electric
Actuator Instruction manual for the procedure to align the valve.
A badly scored rotor will allow "cross talk" between adjacent streams.