Overview
The MAX300
-LG process mass spectrometer is a versatile real-time process analyzer
capable of fully automatic operation. The MAX300
-LG and components are shipped in
an appropriate crate to ensure no damage during shipment. When the instrument arrives,
inventory of all equipment should be taken and any damage or discrepancies reported
both to the freight carrier and Extrel CMS. If no damage is evident, the customer should
prepare the instrument for start up via utility and communication connections discussed
in this manual. The customer then assumes responsibility for any shipping damage.
Obvious physical damage found at installation could void the warranty. As Extrel CMS
does not generally inspect a site prior to installation, it is extremely important to provide
the necessary support for the instrument to avoid significant delays. If a contract with
Extrel CMS is purchased, a service representative may travel onsite to perform a five day
installation. All utility and environmental requirements discussed must be fulfilled prior
to the representative’s arrival.
MAX300
-LG Description
MAX300
-LG has been designed for quantitative analysis of gases and vapors having
molecular weights in the range of 2 to 300 amu. At the heart of the MAX300
-LG is a
quadrupole mass spectrometer that can monitor any practical number of components in a
sample stream. Streams are introduced into the analyzer through an inlet system that is
under control of the data system. The MAX300
-LG major sub systems include the mass
spectrometer hardware, electronic components to operate the spectrometer, vacuum
pumps, heaters, sample inlet valves, a control computer and (optional) communication
modules. Stream sampling valves are mounted on the front of the enclosure.
The MAX300
-LG can be connected to the local area network (LAN) and be accessed
by any PC (client) with access to the LAN. Network access provides all the functions
needed for normal instrument operation. Access is implemented by a web browser
interface with no specialized software required to be installed on the client PC. Access is
password protected and logged for security. Data can be stored in the control computer
and sent to external devices via several communication options.
MAX300
-LG Specifications
Instrument
Dimensions (in/cm) 23.5/59.7 H x 26.5/67.3 W x 19/48.3 D
Weight ~165 lb (75 Kg)
Ambient Temperature 55
o
F to 80
o
F (13°C to 27°C)