EasyManua.ls Logo

Extrel Max300-LG - Page 3

Default Icon
35 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Max300-LG : Description - Specifications - Installation Guide
Extrel
Core Mass Spectrometers
Page 3
Quadrupole Mass Analyzer
• Type Quadrupole
• Rod Size 3/4 inch (19mm)
• Operating Frequency 1.0MHz
• Power Level 200 watt
• Mass Range 2-250 amu
• Scan Speed 1 to 200 amu/sec
• Acquisition Modes Scanning, full profile, selected ion monitoring
(SIM)
• Maximum SIM Ions (No practical limit)
Ion Generation
• Type Electron impact ionization
• Filament Tungsten or Yttrium Oxide coated Iridium
(standard) Others on special request
• Ion Polarity Positive
Ion Detector
• Type Faraday or Faraday-Multiplier (optional)
• Multiplier Type Continuous dynode
• Multiplier Operating Voltage 0 to 3000 V
Vacuum Chamber (Quadrupole Housing)
• Type Single high vacuum pump chamber
• Primary Vacuum Pump 60 liter/second turbomolecular pump
• Secondary Vacuum Pump 1.4 CFM rotary vane pump
• Pressure Monitoring Ion gauge (optional)
Electronics
• Extrel 5221 Card Cage and Control Modules
Control Computer
• Instrumentation Control Windows7 Pro or Windows10 Pro Desktop/Laptop
PC with Questor5
software installed.