Max300-LG : Description - Specifications - Installation Guide
Extrel
Core Mass Spectrometers
Page 4
Minimum PC Requirements:
2 GHz or above CPU
2 GB or above RAM
50 GB free Hard Disk space
RS232 Serial Port for Valve Actuator (Or a USB
2.0 port with a USB to Serial converter)
A USB 2.0 or above port (Or an Ethernet port) for
instrument communication.
MAX300
™
-LG Performance Specifications
• Mass Range 2-250 amu
• Detectable Components Gases or vapors with molecular weights
or ions within the mass range of the instrument.
• Number of Components (No Practical Limit)
• Upper Detection Limit 100%
• Lower Detection Limit 10 parts per million (ppm) with a Faraday
Detector. 10 parts per billion (ppb) with
an electron multiplier.
• Dynamic Range 10
5
with a Faraday detector.
Approximately 10
8
with
a dual Faraday/multiplier detector.
• Stability Better than 0.25% of component’s concentration
(no interference). Example: 2% CO2 will vary
from 1.995% to 2.005%.
• Analysis Repetition Rate The average time required for analysis is
400 milliseconds per component
• Scan Mode MAX300-LG is capable of scanning
the entire mass range
MAX300
™
-LG Environment and Electrical Specifications
• Ambient Temperature Range 55
o
F to 80
o
F
13
o
C to 27
o
C
• Relative Humidity Range 0-90 % ( Non Condensing)