Tokyo Rikakikai Co., Ltd.
CA-1116A 220V
R00
Instruction
Manual
This instruction manual is designed to use the product safely with keeping
its best performance.
Please keep this manual in a place easily accessible to every users.
IMPORTANT
Be sure to read “Safety precautions” before use.
Low Temperature Circulator
Cool Ace

Questions and Answers

  • R
    Richard RandallAug 12, 2025
    What to do if the EYELA A-1116A residual current device cannot be turned on?
    • S
      sharpkellyAug 12, 2025
      If the residual current device cannot be turned on, it could be due to an electric leakage hazard or excess current flow. In the event of an electric leakage hazard, immediately stop operation and contact your local dealer or closest customer service center.
  • A
    Austin EwingAug 15, 2025
    What to do if the EYELA A-1116A refrigeration unit does not work?
    • Y
      ysteeleAug 15, 2025
      If the refrigeration unit isn't working, it could be due to several reasons: the high-pressure switch for the refrigeration unit or overload relay protection circuit might be activated, the refrigeration unit might be impaired, or the water level may be low, exposing the cooling coil. If the thermal load is extremely large, reduce the load. If the refrigeration unit is impaired, stop operation and contact your local dealer or closest customer service center. If the water level is low, fill the circulation liquid in the low-temperature circulation bath.
  • K
    Kevin ArnoldAug 23, 2025
    Why is there no display shown on the indicator when turning on the power switch of EYELA A-1116A Laboratory Equipment?
    • K
      Katherine McculloughAug 23, 2025
      If no display is shown when turning on the power switch, it could be due to several reasons: the mains connector may not be plugged into the outlet completely, power may not be supplied, the residual current device may be turned off or impaired, or the power switch or temperature controller may be impaired. Plug the main connector in after turning off the residual current device and power switch. Turn on the breaker of the distribution board. Turn on the residual current device. If the residual current device is impaired, stop operation and contact your local dealer or closest customer service center.
  • D
    danielriveraAug 27, 2025
    Why is my EYELA A-1116A Laboratory Equipment bath frozen at -7°C or higher?
    • M
      Marie BartonAug 27, 2025
      If the bath internals are frozen, it could be due to poor circulation because of pressure loss of piping, or the temperature controller is broken, or the refrigeration unit cannot be stopped. Open the valve for adjusting discharge volume. Use antifreeze. Using a high-pressure pump is recommended. If the temperature controller is broken or the refrigeration unit cannot be stopped, stop operation and contact your local dealer or closest customer service center.
  • R
    Ryan Jacobson MDAug 31, 2025
    What to do if cool water does not circulate in my EYELA A-1116A?
    • S
      sophia48Aug 31, 2025
      If cool water isn't circulating, it could be that the valve for adjusting discharge volume is closed, dirt adheres to the strainer of the low-temperature circulation bath, or air is being sucked in. Open the valve for adjusting discharge volume. Remove the dirt from the strainer. Remove the drain stopcock and make sure that cool liquid flows, then attach the stopcock again. Turn the pump switch on and off two or three times to release the air.
  • S
    Sean RoblesSep 3, 2025
    What to do if the circulation volume is small in EYELA Laboratory Equipment?
    • R
      robin35Sep 3, 2025
      If the circulation volume is small, it could be because the circulation nozzle has a smaller diameter, the hose is crushed, or the pressure loss of the circulation system is too much. Replace the nozzle with the supplied nozzle that has a larger diameter. Stretch the hose. Reduce the pressure loss.
  • S
    Steven LaneSep 7, 2025
    Why temperature cannot be controlled and there is abnormal indication in EYELA A-1116A Laboratory Equipment?
    • S
      summersbillySep 7, 2025
      If the temperature cannot be controlled and there is an abnormal indication, it is because the temperature controller is in an abnormal state due to noise. Change the installation site. Set the room temperature at 35°C or lower. If that failure mode still remains, stop operation and contact your local dealer or closest customer service center.
  • G
    George SmithNov 18, 2025
    What to do if the EYELA A-1116A Laboratory Equipment alarm is activated?
    • G
      Garrett ClarkNov 18, 2025
      If the alarm is activated on your EYELA Laboratory Equipment, it could be due to several reasons: the high-pressure switch or overload relay holding circuit for the refrigeration unit may be activated, there may be a disconnection of the temperature sensor, an upper or lower temperature limit alarm may have been triggered, or a power failure recovery alarm may have been issued. Set the ambient temperature at 35°C or lower. If the heat load is out of cooling capability, reduce the load. Check the fan for the refrigeration unit. Check the power and voltage, ensuring the rating voltage is 220V±10%. For a disconnected temperature sensor, stop operation and contact your local dealer or closest customer service center. For upper temperature limit alarm, after confirming that the bath temperature...

Summary

Safety Information

Understanding Warning Signals

Explains hazard symbols, signal words, and their meanings for safe operation.

Product Overview

Technical Specifications

Detailed technical data including dimensions, temperature range, cooling capability, and power.

Controls and Alarms

Control Panel Functions

Description of buttons, indicators, and the display on the unit's control panel.

Safety Alarm Features

Explanation of safety devices, alarm functions, and their causes.

Alarm Notifications

Details various alarm types, their visual indicators, and potential causes.

Installation and Setup

Operation Guide

Advanced Settings and Troubleshooting

Troubleshooting Guide

Solutions for common problems, error indications, and alarm troubleshooting.

EYELA A-1116A Specifications

General IconGeneral
BrandEYELA
ModelA-1116A
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals