Tokyo Rikakikai Co., Ltd.
CA-1116A 220V
R00
Instruction
Manual
This instruction manual is designed to use the product safely with keeping
its best performance.
Please keep this manual in a place easily accessible to every users.
IMPORTANT
Be sure to read “Safety precautions” before use.
Low Temperature Circulator
Cool Ace

Questions and Answers

  • N
    Nicholas LewisAug 14, 2025
    What to do if the EYELA CA-1116A Laboratory Equipment residual current device cannot be turned on?
    • R
      Roy MillerAug 14, 2025
      If the residual current device cannot be turned on, it could be due to an electric leakage hazard or excess current flow. In the event of an electric leakage hazard, immediately stop operation and contact your local dealer or closest customer service center.
  • E
    Emily GarnerAug 17, 2025
    What to do if the EYELA CA-1116A Laboratory Equipment refrigeration unit does not work?
    • M
      Michael TorresAug 17, 2025
      If the refrigeration unit isn't working, it could be due to several reasons: the high-pressure switch for the refrigeration unit or overload relay protection circuit might be activated, the refrigeration unit might be impaired, or the water level may be low, exposing the cooling coil. If the thermal load is extremely large, reduce the load. If the refrigeration unit is impaired, stop operation and contact your local dealer or closest customer service center. If the water level is low, fill the circulation liquid in the low-temperature circulation bath.
  • M
    Misty MillerAug 25, 2025
    Why is there no display shown on the indicator when turning on the power switch of EYELA CA-1116A?
    • K
      Katherine SilvaAug 26, 2025
      If no display is shown when turning on the power switch, it could be due to several reasons: the mains connector may not be plugged into the outlet completely, power may not be supplied, the residual current device may be turned off or impaired, or the power switch or temperature controller may be impaired. Plug the main connector in after turning off the residual current device and power switch. Turn on the breaker of the distribution board. Turn on the residual current device. If the residual current device is impaired, stop operation and contact your local dealer or closest customer service center.
  • S
    Scott CochranAug 29, 2025
    Why is my EYELA CA-1116A Laboratory Equipment bath frozen at -7°C or higher?
    • R
      Robert WilliamsAug 29, 2025
      If the bath internals are frozen, it could be due to poor circulation because of pressure loss of piping, or the temperature controller is broken, or the refrigeration unit cannot be stopped. Open the valve for adjusting discharge volume. Use antifreeze. Using a high-pressure pump is recommended. If the temperature controller is broken or the refrigeration unit cannot be stopped, stop operation and contact your local dealer or closest customer service center.
  • Z
    Zachary GillSep 1, 2025
    What to do if cool water does not circulate in my EYELA Laboratory Equipment?
    • B
      bchaseSep 1, 2025
      If cool water isn't circulating, it could be that the valve for adjusting discharge volume is closed, dirt adheres to the strainer of the low-temperature circulation bath, or air is being sucked in. Open the valve for adjusting discharge volume. Remove the dirt from the strainer. Remove the drain stopcock and make sure that cool liquid flows, then attach the stopcock again. Turn the pump switch on and off two or three times to release the air.
  • A
    angelica03Sep 5, 2025
    What to do if the circulation volume is small in EYELA CA-1116A?
    • M
      michaelgarciaSep 5, 2025
      If the circulation volume is small, it could be because the circulation nozzle has a smaller diameter, the hose is crushed, or the pressure loss of the circulation system is too much. Replace the nozzle with the supplied nozzle that has a larger diameter. Stretch the hose. Reduce the pressure loss.
  • P
    penadianaSep 9, 2025
    Why temperature cannot be controlled and there is abnormal indication in EYELA CA-1116A?
    • S
      Samantha VillanuevaSep 9, 2025
      If the temperature cannot be controlled and there is an abnormal indication, it is because the temperature controller is in an abnormal state due to noise. Change the installation site. Set the room temperature at 35°C or lower. If that failure mode still remains, stop operation and contact your local dealer or closest customer service center.
  • V
    vbutlerNov 20, 2025
    What to do if the EYELA CA-1116A alarm is activated?
    • N
      Natalie ReedNov 20, 2025
      If the alarm is activated on your EYELA Laboratory Equipment, it could be due to several reasons: the high-pressure switch or overload relay holding circuit for the refrigeration unit may be activated, there may be a disconnection of the temperature sensor, an upper or lower temperature limit alarm may have been triggered, or a power failure recovery alarm may have been issued. Set the ambient temperature at 35°C or lower. If the heat load is out of cooling capability, reduce the load. Check the fan for the refrigeration unit. Check the power and voltage, ensuring the rating voltage is 220V±10%. For a disconnected temperature sensor, stop operation and contact your local dealer or closest customer service center. For upper temperature limit alarm, after confirming that the bath temperature...

Summary

Important Safety Information

Safe Operation of the Unit

Key Safety Precautions

Covers handling flammable solvents, moving the unit, and contact with cooling fins.

Operating Portion and Control Panel

Control Panel Functions

Explains the layout and functions of the control panel buttons and indicators.

Safety and Alarm Features

Details the built-in safety devices and alarm mechanisms of the unit.

Installation

Installation Environment

Specifies ideal site conditions for installation, focusing on ventilation and temperature.

Utility Connection

Explains how to safely connect the unit to a power source, including grounding.

Operation

Preparation

Covers essential steps before operation, like connecting hoses and filling the bath.

How to Operate the Unit

Describes the initial power-on sequence and basic operation.

Setting the Temperature

Explains how to set the desired operating temperature for the unit.

Starting Operation

Details the steps to start the circulation pump and the refrigeration unit.

Stopping Operation

Guides on how to safely stop the unit and its circulation pump.

Troubleshooting

Trouble and Countermeasure

Provides solutions for common problems and error indications.

Maintenance and Checkup

Cleaning and Caring the Product

Provides instructions for cleaning the unit's components and general care.

Disposal of the Product

EYELA CA-1116A Specifications

General IconGeneral
BrandEYELA
ModelCA-1116A
CategoryLaboratory Equipment
LanguageEnglish