D
Debra PattersonAug 21, 2025
Why is my EYELA Accessories unit not cooling down properly?
- RRaymond GarzaAug 21, 2025
If your EYELA Accessories unit isn't cooling down well, check the setting temperature to ensure it's appropriate. If the ambient temperature is too high, set the value at 30°C (LTI-400E) or lower than 35°C. Also, too many samples and vessels in the chamber can prevent convection flow of heat, so reduce their number. If the refrigeration unit is frosted, check the condition of frost formation and execute manual forced Defrosting. In case the refrigeration unit does not work, immediately stop running the product and contact your local agency or the closest customer service center.