What causes the over heat alarm to activate on my EYELA LTI- 400E Accessories?
- SSonya BrownSep 7, 2025
The over heat alarm activates on your EYELA Accessories if the setting value for the over heat protector is too low. For the settings for over heat protector, set the temperature +10° higher than the setting temperature of the temperature controller. If the setting temperature is too high, set the value up to 30°.