14
In order to update the Instrument Software with a New Firmware Release
follow the suggested procedures:
TO UPDATE INSTALLED UNITS
- Download the interface software “FLS Calibration System” and the Updated
Software on www.snet.it
- Launch the software "FLS Calibration System" on the laptop
- Select OPTION and then UPGRADE FIRMWARE
- Conrm the "Firmware Upgrade" procedure by ENTER
- Connect M9.50 to the laptop by the USB cable
- Select the item (M9.XX) which appears on the "Navigation" area on the "FLS
Calibration System" software
- Conrm FW UPGRADE and select the Updated Software
NOTE: At the end of the procedure restart the instruments in order to refresh
M9.50 software (It takes 90 seconds to refresh the SW. Please do not interrupt
the restarting process).
TO UPDATE NEW UNITS
- Download the interface software “FLS Calibration System” and the updated
software on www.snet.it.
- Launch the software "FLS Calibration System" on the laptop
- Push together ENTER and ESC powering the monitor
- Connect M9.50 to the laptop by the USB cable
- Select the item (M9.XX) which appears on the "Navigation" area on the
software
"FLS Calibration System"
- Conrm FW UPGRADE and select the Updated Software
NOTE: At the end of the procedure restart the instruments in order to refresh
M9.50 software (It takes 90 seconds to refresh the SW. Please do not interrupt
the restarting process).
SOFTWARE UPDATING