Table 6. Hazardous Area Classifications for Fisher 582i Converter
(1)
—ATEX
Certificate Certification Obtained Entity Rating Temperature Code
ATEX
II 1 G & D
Intrinsically Safe
Gas
Ex ia IIC T4/T5/T6 Ga
Ui = 30 VDC
Ii = 150 mA
Pi = 1.25 W
Ci = 0 nF
Li = 0 mH
T4 (Tamb ≤ 71°C)
T5 (Tamb ≤ 62°C)
T6 (Tamb ≤ 47°C)
Dust
Ex ia IIIC Da T109 °C (Tamb ≤ 71°C) / T100 °C (Tamb v 62°C)
/ T85 °C (Tamb ≤ 47°C)
- - -
II 2 G & D
Flameproof
Gas
Ex d IIC T6 Gb
- - - T6 (Tamb ≤ 71°C)
Dust
Ex tb IIIC T74 °C Db (Tamb ≤ 71°C)
- - -
II 3 G & D
Type n
Gas
Ex nA IIC T6 Gc
- - -
T6 (Tamb ≤ 71°C)
Dust
Ex tc IIIC Dc T85°C (Tamb ≤ 71°C)
- - -
1.These hazardous area classification also apply to 3582i positioners.