EasyManua.ls Logo

Fisher 582i - IECEX

Fisher 582i
56 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Instruction Manual
D200138X012
3582, 582i, and 3583
February 2015
12
IECEx
Conditions of Certification
Intrinsically Safe
WARNING
Substitution of components may impair intrinsic safety.
-40_C Ta +71_C; T6 (Ta +47_C); T5 (Ta +62_C); T4 (Ta +71_C)
Entity Parameters
Ui = 30 V, li = 150 mA, Pi = 1.25 W, Ci = 0 nF, Li = 0 mH
Flameproof
WARNING
Disconnect power before opening.
-40_C Ta +71_C; T6 (Ta +71_C)
Type n
WARNING
Disconnect power before opening.
-40_C Ta +71_C; T6 (Ta +71_C)
Refer to table 7 for approval information.
Table 7. Hazardous Area Classifications for Fisher 582i Converter
(1)
—IECEx
Certificate Certification Obtained Entity Rating Temperature Code
IECEx
Intrinsically Safe
Gas
Ex ia IIC T4/T5/T6 Ga
Ui = 30 VDC
Ii = 150 mA
Pi = 1.25 W
Ci = 0 nF
Li = 0 mH
T4 (Tamb 71°C)
T5 (Tamb 62°C)
T6 (Tamb 47°C)
Flameproof
Gas
Ex d IIC T6 Gb
- - - T6 (Tamb 71°C)
Type n
Gas
Ex nA IIC T6 Gc
- - - T6 (Tamb 71°C)
1.These hazardous area classification also apply to 3582i positioners.

Related product manuals