Intended Use
The ProVu 3.5” Display is a digital display intended to show live images from the ProVu Video Laryngoscope (VL)
only, via a micro-USB connection, as an aid to intubation for the hospital use.
Product Description
The ProVu 3.5” Display is a non-sterile, reusable digital display intended to show live images from the ProVu Video
Laryngoscope.
When connected to the ProVu device, a live video is shown on the display, which can be used to further aid
intubation of the patient.
The display can be recharged on the charging dock provided or with the ProVu charging cable and, when fully
charged, will provide a minimum of 1 hour of continuous use.
Flexicare does not recommend the modication of the display. Any attempts to do so, will result in the failure of
the display and voids warranty.
Precautions
- Ensure that you always have easy access to the power cord, plug or adapter to disconnect this display from
the power.
- Inspect the ProVu 3.5” Display and all accessories for damage prior to each use. Do not use if it appears
damaged.
- Charge the ProVu 3.5” Display fully before use. For shipping requirements, ProVu 3.5” Displays are shipped
with zero charge.
- Before charging, ensure that all components are completely dry, including the micro-USB port.
- If not connected to ProVu VL and left unmoved for more than 5 minutes, the ProVu 3.5” Display will switch into
“sleep mode” and the screen will go blank. This is normal. To reverse this, briey press the Power Button. A
prompt image should be seen after 2 seconds.
Warnings & Cautions
- ProVu 3.5” Display and Accessories should only be used by trained medical professionals.
- Do not use within Magnetic Resonance Imaging (MR/MRI) environments or with other high interference
sources.
- Do not sterilize the display.
- The use of devices and other accessories which are not approved by Flexicare may place the patient at risk. It is
recommended that the ProVu 3.5” Display should only be used with the ProVu Video Laryngoscope and ProVu
accessories.
- Always disconnect from ProVu 3.5” Display from the charging dock before cleaning.
- Ensure the voltage recommendations of the display wall plug are compatible with power supplies.
- Do not clean the display or display accessories using any abrasive agents or mineral acids.
- Do not use sterilants with caustic ingredients, such as surgical scrub solutions, peroxide solutions, bleaches,
or povidone-iodine solutions.
- Inadequate battery levels may result in sudden failure without warning. Check sucient battery level before
use.
- Do not disassemble display.
- Do not modify the display.
- Battery is non-replaceable and non-removable.
- There are no user serviceable parts in the device. Any faulty device should be replaced or returned within
warranty to the manufacturer.
- Do not attempt to charge while battery is in use.
- Do not attempt maintenance of the device during use.
- Do not submerge.
- If the display screen does not illuminate from “sleep mode”, it is likely the battery needs to be charged or the
device needs to be reset. Follow Troubleshooting steps below for proper steps.