6.3 Thermal Considerations
It is important to minimize any temperature gradient across the camera. The sensor should be mounted in
such a
fashion so as to isolate it from heat loads such as electronics, heaters, and non-symmetric external
heating. The
surrounding area must be able to support and withstand the dissipation of up to 160 mW of heat by the
camera.
6.4 Optical Considerations
The optical keep-out zone is described by the three-dimensional field of view cone within the Lepton with
Radiometry STEP file. To avoid mechanical vignetting, do not impinge upon the keep-out zone defined by this
cone.
7 Image Characteristics
The information given in Table 13 applies across the full operating temperature range.
Table 13 - Image Characteristics
Number of adjacent defective
pixels
“Adjacent” means any of the 8
nearest neighbors (or
nearest
5 for an edge pixel, nearest 3
for a corner).
Note(s)
1. Scene dynamic range is a function of sensor characteristics and ambient temperature. Range values reported are typical values at
room temperature ambient. See Table 1- Key Specifications for details.
2. Only single-pixel defects are allowed (no clusters).