17 VT8-600_1000-en-US_AA 10/18
7. Specifications
7.1 Electrical Specifications
Accuracy is given as ± (% of reading + least significant digits) at 73.4
F ±9
F (23
C ±5
C) with
relative humidity <80%. Accuracy is specified for a period of one year after calibration.
Function
Range
Resolution
Accuracy
(reading)
AC Current
100A (VT8-600)
200A (VT8-1000)
0.1A
± (2.5% + 5 digits)
True RMS, Frequency Response 50 ~ 60Hz
DC Current
100A (VT8-600)
200A (VT8-1000)
0.1A
± (2.5% + 5 digits)
Use the DC Zero function to null the display reading
AC Voltage
6.000V
0.001V
± (1.2% + 5 digits)
60.00V
0.01V
± (1.2% + 3 digits)
600.0V
0.1V
± (1.5% + 2 digits)
1000V (VT8-1000)
0.1V
± (1.5% + 3 digits)
True RMS voltage applicable to 10% ~ 100% of the range
Input Impedance: ≥ 10MΩ; Frequency response 45 ~ 66Hz
DC Voltage
6.000V
0.001V
± (0.8% + 3 digits)
60.00V
0.01V
600.0V
0.1V
± (1.0% + 2 digits)
1000V (VT8-1000)
0.1V
± (1.0% + 3 digits)
Input Impedance: ≥ 10MΩ
Resistance
600.0
0.1
± (1.2% + 2 digits)
6.000k
0.001k
± (1.0% + 2 digits)
60.00k
0.01k
600.0k
0.1k
6.000M
0.001M
± (1.2% + 2 digits)
60.00M
0.01M
± (1.5% + 5 digits)
Continuity
600.0
0.1
-----
< 10 beeper sounds. 10 to 100 undetermined. > 100 beeper off (OL
displayed). Open circuit voltage approx. 1.2V