17 / ENGLISH ENGLISH / 18
If you have any questions regarding Flow,
don’t hesitate to contact customer support at
support@flowneuroscience.com
Flow Neuroscience AB
Bredgatan 4
211 30 Malmö
Skåne Län, Sweden
16. Contact Us
EMISSION LIMITS & MEASUREMENT
PHENOMENON STANDARD TEST METHOD TEST LEVELS RESILT
RADIATED RF
EMISSION
IEC 60601-1-2
CISPR 11
ETSI EN 301
489-1
30~1000 MHz ; 1~3 GHz &
3~6 GHz
(Measurement Uncertainty
4.24 dB)
PASS
CONDUCTED
RF EMISSION
IEC 60601-1-2
CISPR 11
ETSI EN 301
489-1
0,15~30 MHz
(Measurement Uncertainty
3.35 dB)
PASS
VOLTAGE
FLUCTUATION
& FLICKER
IEC 60601-1-2 IEC 61000-3-3
EUT VALUES & LIMITS
Pst: 0.028 ; Limit 1.00
Plt: 0.028 ; Limit 0.65
dc [%]: 0.0000 ; Limit 3.30
dmax [%]: 0.053 ; Limit 4.00
dt [s]: 0.000 ; Limit 0.50
PASS
17. EMC - Test Results
IMMUNITY MEASUREMENT
PHENOMENON STANDARD
TEST
METHOD
TEST LEVELS RESILT
ELECTROSTATIC
DISCHARGE
IEC
60601-1-2
IEC 61000-
4-2
±2 kV, ±4 kV, ± 6 kV, ± 8 kV -
Contact Discharge
± 2 kV, ± 4 kV, ± 8 kV, ± 15
kV - Air Discharge
PASS
RF
ELECTROMAGNETIC
FIELDS
IEC
60601-1-2
IEC 61000-
4-3
10 V/m
80 – 2700 MHz
80 % AM at 1 kHz
PASS
EL FAST TRANSIENTS
COMMON MODE /
BURSTS
IEC
60601-1-2
IEC 61000-
4-4
± 2 kV for A.C. Power Lines
100 kHz Burst Frequency
PASS
SURGES LINE TO LINE
IEC
60601-1-2
IEC 61000-
4-5
± 0.5 kV & ±1 kV At 0°,
90°, 180°, 270° ; Repetition
Rate: 60 S
PASS
SURGES LINE
TO GROUND
IEC
60601-1-2
IEC 61000-
4-5
± 0.5 kV, ±1 & ±2 kV At 0°,
90°, 180°, 270°
PASS
RF COMMON MODE
IEC
60601-1-2
IEC 61000-
4-6
3 & 6 V for A.C. Main Port
; 80%, 1 kHz Amplitude
Modulation)
Frequency: 150kHz to 80MHz
PASS
VOLTAGE DIPS &
INTERRUPTIONS
IEC
60601-1-2
IEC 61000-
4-11
0 % UT; 0,5 Cycle At 0°,
45°,90°, 135°,180°, 225°,
270°, 315°,
0 % UT; 1.0 Cycle
70 % UT; 25/30 Cycles
PASS