EasyManua.ls Logo

Flow Neuroscience FLOW - Page 19

Flow Neuroscience FLOW
23 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
33 / SVENSKA SVENSKA / 34
16. Kontakt
Om du har några frågor angående Flow,
tveka inte att kontakta kundsupport på
support@flowneuroscience.com
Flow Neuroscience AB
Bredgatan 4
211 30 Malmö
Skåne Län, Sweden
IMMUNITY MEASUREMENT
PHENOMENON STANDARD
TEST
METHOD
TEST LEVELS RESILT
ELECTROSTATIC
DISCHARGE
IEC
60601-1-2
IEC 61000-
4-2
±2 kV, ±4 kV, ± 6 kV, ± 8 kV -
Contact Discharge
± 2 kV, ± 4 kV, ± 8 kV, ± 15
kV - Air Discharge
PASS
RF
ELECTROMAGNETIC
FIELDS
IEC
60601-1-2
IEC 61000-
4-3
10 V/m
80 – 2700 MHz
80 % AM at 1 kHz
PASS
EL FAST TRANSIENTS
COMMON MODE /
BURSTS
IEC
60601-1-2
IEC 61000-
4-4
± 2 kV for A.C. Power Lines
100 kHz Burst Frequency
PASS
SURGES LINE TO LINE
IEC
60601-1-2
IEC 61000-
4-5
± 0.5 kV & ±1 kV At 0°,
90°, 180°, 270° ; Repetition
Rate: 60 S
PASS
SURGES LINE
TO GROUND
IEC
60601-1-2
IEC 61000-
4-5
± 0.5 kV, ±1 & ±2 kV At 0°,
90°, 180°, 270°
PASS
RF COMMON MODE
IEC
60601-1-2
IEC 61000-
4-6
3 & 6 V for A.C. Main Port
; 80%, 1 kHz Amplitude
Modulation)
Frequency: 150kHz to 80MHz
PASS
VOLTAGE DIPS &
INTERRUPTIONS
IEC
60601-1-2
IEC 61000-
4-11
0 % UT; 0,5 Cycle At 0°,
45°,90°, 135°,180°, 225°,
270°, 315°,
0 % UT; 1.0 Cycle
70 % UT; 25/30 Cycles
PASS
EMISSION LIMITS & MEASUREMENT
PHENOMENON STANDARD TEST METHOD TEST LEVELS RESILT
RADIATED RF
EMISSION
IEC 60601-1-2
CISPR 11
ETSI EN 301
489-1
30~1000 MHz ; 1~3 GHz &
3~6 GHz
(Measurement Uncertainty
4.24 dB)
PASS
CONDUCTED
RF EMISSION
IEC 60601-1-2
CISPR 11
ETSI EN 301
489-1
0,15~30 MHz
(Measurement Uncertainty
3.35 dB)
PASS
VOLTAGE
FLUCTUATION
& FLICKER
IEC 60601-1-2 IEC 61000-3-3
EUT VALUES & LIMITS
Pst: 0.028 ; Limit 1.00
Plt: 0.028 ; Limit 0.65
dc [%]: 0.0000 ; Limit 3.30
dmax [%]: 0.053 ; Limit 4.00
dt [s]: 0.000 ; Limit 0.50
PASS
17. EMC - Testresultat